Mouse Complement factor D enzyme-linked immunoassay kit(one step)
| Specification | 96 Test |
|---|---|
| Sensitivity | 0.09ng/ml (10 μl) |
| Standard Curve Range | 0.69 ~500 ng/ml |
| Standard Curve Gradient | 7 Points/3 Folds |
| Number of Incubations | 1 |
| Detectable sample | serum, plasma |
| Sample Volume | 10 μl |
| Type | Fully Ready-to-Use |
| Operation Duration | 60min |
| ng/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0049 | 0.0061 | 0.0055 | |
| 0.69 | 0.0204 | 0.0224 | 0.0214 | 0.0159 |
| 2.06 | 0.0521 | 0.0520 | 0.0521 | 0.0466 |
| 6.17 | 0.1302 | 0.1442 | 0.1372 | 0.1317 |
| 18.52 | 0.4134 | 0.4098 | 0.4116 | 0.4061 |
| 55.56 | 0.9880 | 1.0940 | 1.0410 | 1.0355 |
| 166.67 | 2.3720 | 2.4590 | 2.4155 | 2.4100 |
| 500.00 | 4.1428 | 3.9500 | 4.0464 | 4.0409 |
Precision
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(ng/ml) | 12.4 | 58.0 | 171.5 | 10.9 | 52.6 | 158.0 |
| Standard Deviation | 0.8 | 4.4 | 13.0 | 0.6 | 3.0 | 10.2 |
| Coefficient of Variation(%) | 6.8 | 7.6 | 7.6 | 5.8 | 5.8 | 6.5 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of Mouse Complement factor D into health mouse serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 94% to 119% with an overall mean recovery of 115%.
Sample Values
| Sample Matrix | Sample Evaluated | Range (μg/ml) | Detectable (%) | Mean of Detectable (μg/ml) |
|---|---|---|---|---|
| Serum | 30 | 13.55-19.91 | 100 | 17 |
Serum/Plasma – Thirty samples from apparently healthy mice were evaluated for the presence of Mouse Complement factor D in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
Product Data Sheet
Background: Complement factor D
Complement Factor D (also Adipsin, C3 convertase activator and Properdin factor D) is a 37‑43 kDa member of the S1 peptidase family of enzymes. In mouse, it is secreted by fat cells and helps form C3 convertase, a complex of C3b, Factor B and Factor D. This complex cleaves C3 into C3b and C3a, the latter of which is converted into ASP, a molecule that binds the adipocyte receptor C5L2 and induces triglyceride synthesis. Mature mouse Factor D is 234 amino acids (aa) in length (aa 26‑259). In contains one peptidase S1 domain that cleaves an Arg-Lys bond (aa 26‑254). There are at least two potential alternative splice forms. One shows a two aa substitution for aa 1‑71, and a second shows a 15 aa substitution for aa 1‑107. Mature mouse factor D shares 85% and 68% aa identity with rat and human Factor D, respectively.