Human ICAM-1/CD54 enzyme-linked immunoassay kit

CAT: EHY0008 Datasheet
Specification 96 Test
Sensitivity 1.22 pg/ml (50 μl);0.05 ng/ml (10 μl)
Standard Curve Range 0.14~100 ng/ml
Standard Curve Gradient 7 Points/3 Folds
Number of Incubations 2
Sample Volume 50 μl/10 μl
Type Fully Ready-to-Use
Operation Duration 75min
ng/ml O.D. Average Corrected
0.00 0.0050 0.0046 0.0048
0.14 0.0210 0.0195 0.0203 0.0155
0.41 0.0519 0.0540 0.0530 0.0482
1.23 0.1489 0.1465 0.1477 0.1429
3.70 0.4238 0.4106 0.4172 0.4124
11.11 1.0740 1.0970 1.0855 1.0807
33.33 2.3350 2.3240 2.3295 2.3247
100.00 3.1570 3.3020 3.2295 3.2247

Precision

Intra-assay Precision Inter-assay Precision
Sample Number S1 S2 S3 S1 S2 S3
22 22 22 6 6 6
Average(ng/ml) 1.8 9.0 32.7 1.7 8.7 31.0
Standard Deviation 0.1 0.4 2.2 0.1 0.4 1.0
Coefficient of Variation(%) 3.3 4.8 6.8 5.8 4.3 3.1

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery

The spike recovery was evaluated by spiking 3 levels of human ICAM-1/CD54 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 83% to 120% with an overall mean recovery of 105%.

Sample Values

Sample Matrix Sample Evaluated Range (ng/ml) Detectable (%) Mean of Detectable (ng/ml)
Serum 30 55.22-176.34 100.0 103.75

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of ICAM-1/CD54 in this assay. No medical histories were available for the donors.

Background: ICAM-1/CD54

Intercellular Adhesion Molecule-1 (ICAM-1, CD54) binds the leukocyte integrins LFA-1 and Mac-1. ICAM-1 expression is weak on leukocytes, epithelial and resting endothelial cells, as well as some other cell types, but expression can be stimulated by IFN-gamma, TNF-alpha, IL-1 beta and LPS. Mouse and human ICAM-1 share approximately 54% amino acid identity.

Soluble ICAM-1 is found in a biologically active form in serum, probably as a result of proteolytic cleavage from the cell surface, and is elevated in patients with various inflammatory syndromes such as septic shock, leukocyte adhesion deficiency syndrome (LAD), cancer and transplantation.