Mammalian CD62P/P-Selectin enzyme-linked immunoassay kit

CAT: EH0289 Datasheet
Specification 96 Test
Sensitivity 8.19 pg/ml (50 μl);70.99 pg/ml (10 μl);
Standard Curve Range 156.25~10000 pg/ml
Standard Curve Gradient 7 Points
Number of Incubations 2
Detectable sample Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc.
Sample Volume 50 μl/10 μl
Type Fully Ready-to-Use
Operation Duration 120min
pg/ml O.D. Average Corrected
0.00 0.0444 0.0441 0.0443
156.25 0.1587 0.1492 0.1540 0.1097
312.50 0.2757 0.2684 0.2721 0.2278
625.00 0.4935 0.4962 0.4949 0.4506
1250.00 0.8932 0.8620 0.8776 0.8334
2500.00 1.4950 1.5650 1.5300 1.4858
5000.00 2.3990 2.4770 2.4380 2.3938
10000.00 3.3780 3.4450 3.4115 3.3673

Precision

Intra-assay Precision Inter-assay Precision
Sample Number S1 S2 S3 S1 S2 S3
22 22 22 6 6 6
Average(pg/ml) 199.2 1022.8 3359.7 161.6 695.0 2182.5
Standard Deviation 10.0 44.6 381.0 7.6 29.3 104.0
Coefficient of Variation(%) 5.0 4.4 11.3 4.7 4.2 4.8

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery

The spike recovery was evaluated by spiking 3 levels of mammalian CD62P/P-Selectin into health mammalian serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 80% to 115% with an overall mean recovery of 94%.

Sample Values

Sample Matrix Sample Evaluated Range (ng/ml) Detectable (%) Mean of Detectable (ng/ml)
Human Serum307.85-29.1410018.17
Mouse Serum30355.11-903.67100572.82
Rat Serum3026.18-72.1210046.71
Monkey Serum14.321004.32

Serum/Plasma – Thirty samples from apparently healthy volunteers, thirty samples from apparently healthy mice, thirty samples from apparently healthy rats and one sample from healthy monkeys were evaluated for the presence of mammalian CD62P/P-Selectin in this assay. No medical histories were available for the donors.
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.

Background: CD62P/P-Selectin

E-Selectin and P-Selectin are related (44% amino acid sequence identity) adhesion molecules containing multiple short complement-like repeats and an N-terminal C‑type lectin domain in their extracellular regions. E-Selectin is expressed on activated vascular endothelial cells and P-Selectin occurs on activated platelets and endothelial cells.