Human TRAF1 enzyme-linked immunoassay kit

CAT: EH0215 Datasheet
Specification 96 Test
Sensitivity 2.85 pg/ml (50 μl);7.62 pg/ml (10 μl);
Standard Curve Range 20.48~5000 pg/ml
Standard Curve Gradient 7 Points
Number of Incubations 2
Detectable sample Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc.
Sample Volume 50 μl/10 μl
Type Ready-to-Use
Operation Duration 120min
pg/ml O.D. Average Corrected
0.00 0.0352 0.0374 0.0363
20.48 0.0482 0.0512 0.0497 0.0134
51.20 0.0685 0.0666 0.0676 0.0313
128.00 0.1148 0.1100 0.1124 0.0761
320.00 0.2349 0.2201 0.2275 0.1912
800.00 0.5329 0.5233 0.5281 0.4918
2000.00 1.2500 1.2230 1.2365 1.2002
5000.00 2.8350 2.7000 2.7675 2.7312

Precision

Intra-assay Precision Inter-assay Precision
Sample Number S1 S2 S3 S1 S2 S3
22 22 22 6 6 6
Average() 89.7 437.9 1342.2 79.1 414.5 1371.4
Standard Deviation 4.9 16.7 58.5 5.9 8.8 47.9
Coefficient of Variation(%) 5.5 3.8 4.4 7.4 2.1 3.5

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery

The spike recovery was evaluated by spiking 3 levels of human TRAF1 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from72% to 94% with an overall mean recovery of 84%.

Sample Values

Sample Matrix Sample Evaluated Range () Detectable (%) Mean of Detectable ()
Serum 30 n.d.-43.04 63 14.45

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of TRAF1 in this assay. No medical histories were available for the donors.

Background: TRAF1

TRAF1(肿瘤坏死因子受体相关因子1)是一种调控炎症与细胞凋亡的细胞内衔接蛋白