Human ICAM-3/CD50 enzyme-linked immunoassay kit
| Specification | 96 Test |
|---|---|
| Sensitivity | 7.09pg/ml (50 μl);15.79 pg/ml (10 μl); |
| Standard Curve Range | 40.96 ~10000 pg/ml |
| Standard Curve Gradient | 7 Points |
| Number of Incubations | 2 |
| Detectable sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
| Sample Volume | 50 μl/10 μl |
| Type | Fully Ready-to-Use |
| Operation Duration | 120min |
| pg/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0388 | 0.0398 | 0.0393 | |
| 40.96 | 0.0604 | 0.0594 | 0.0599 | 0.0206 |
| 102.40 | 0.0958 | 0.0936 | 0.0947 | 0.0554 |
| 256.00 | 0.1866 | 0.1878 | 0.1872 | 0.1479 |
| 640.00 | 0.3785 | 0.3978 | 0.3882 | 0.3489 |
| 1600.00 | 0.8924 | 0.8582 | 0.8753 | 0.8360 |
| 4000.00 | 1.9740 | 1.9330 | 1.9535 | 1.9142 |
| 10000.00 | 3.5690 | 3.5980 | 3.5835 | 3.5442 |
Precision
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 305.0 | 1111.8 | 3290.7 | 230.0 | 1081.0 | 3271.7 |
| Standard Deviation | 5.7 | 29.7 | 61.8 | 10.0 | 51.1 | 86.0 |
| Coefficient of Variation(%) | 1.9 | 2.7 | 1.9 | 4.4 | 4.7 | 2.6 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human ICAM-3/CD50 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 98% to 114% with an overall mean recovery of 103%.
Sample Values
| Sample Matrix | Sample Evaluated | Range (ng/ml) | Detectable (%) | Mean of Detectable (ng/ml) |
|---|---|---|---|---|
| Serum | 30 | 93.54-214.96 | 100 | 143.28 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of human ICAM-3/CD50 in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
Product Data Sheet
Background: ICAM-3/CD50
All ICAM proteins are type I transmembrane glycoproteins, contain 2-9 immunoglobulin-like C2-type domains, and bind to the leukocyte adhesion LFA-1 protein. This protein is constitutively and abundantly expressed by all leucocytes and may be the most important ligand for LFA-1 in the initiation of the immune response. It functions not only as an adhesion molecule, but also as a potent signalling molecule. ICAM 3 is the major ligand for LFA-1 in the resting state.