Rat CRP/C-Reactive Protein ELISA Kit检测试剂盒
规格 | 96 Test |
---|---|
灵敏度 | 27.24 pg/ml (50 μl);63.91 pg/ml (10 μl) |
标准曲线范围 | 81.92~20000 pg/ml |
标准曲线梯度 | 7点 |
孵育次数 | 2 |
加样体积 | 50 μl/10 μl |
类型 | 完全即用型 |
操作时长 | 120min |
pg/ml | O.D. | Average | Corrected | |
---|---|---|---|---|
0.00 | 0.0248 | 0.0257 | 0.0253 | |
81.92 | 0.0392 | 0.0371 | 0.0382 | 0.0232 |
204.80 | 0.0777 | 0.0792 | 0.0785 | 0.0635 |
512.00 | 0.1553 | 0.1399 | 0.1476 | 0.1326 |
1280.00 | 0.3510 | 0.3545 | 0.3528 | 0.3378 |
3200.00 | 0.9028 | 0.8789 | 0.8909 | 0.8759 |
8000.00 | 2.0410 | 2.1280 | 2.0845 | 2.0695 |
20000.00 | 3.6460 | 3.9210 | 3.7835 | 3.7685 |
精度
Intra-assay Precision | Inter-assay Precision | |||||
Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
22 | 22 | 22 | 6 | 6 | 6 | |
Average(pg/ml) | 486.6 | 2033.5 | 6246.1 | 483.8 | 2222.2 | 6448.2 |
Standard Deviation | 25.5 | 121.9 | 330.3 | 15.4 | 112.6 | 123.97 |
Coefficient of Variation(%) | 5.2 | 6.0 | 4.8 | 3.2 | 5.1 | 1.9 |
板内精密度:3 个已知浓度的样本酶标板内重复测定 20 次,评估酶标板内的精密度。
板间精密度:3 个已知浓度的样本酶标板间重复检测 6 次,评估酶标板间的精密度。
加标回收
健康人血清加入 3 个不同浓度水平的大鼠CRP,未加大鼠CRP 的血清作为本底,计算回收率。回收率的范围从 74 % 至 137 %,平均回收率为 103 %。
样本值
Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
---|---|---|---|---|
Serum | 30 | 96.31-1708.35 | 100 | 398.28 |
应用本试剂盒,检测 30 个健康大鼠的血清样本,大鼠的用药史不详。n.d. = 测不到浓度值。样本的浓度值低于灵敏度被认为测不到浓度值。
产品数据表
背景: CRP/C-Reactive Protein
C反应蛋白(CRP),也称为Pentraxin 1,是一种分泌的五聚体蛋白,可作为先天免疫反应的传感器和激活剂。在人类中,它是一种主要的急性期蛋白质;其循环浓度在炎症发作时急剧升高。然而,在小鼠中,血清CRP水平在炎症期间仅略有增加,并且类似的急性期作用由Pentraxin 2填充。CRP结合,调理和诱导细菌和凋亡细胞的吞噬作用。它通过结合补体级联中的几种蛋白质以及巨噬细胞和树突状细胞上的 Fc γ RI、Fc γ RIIA 和 Fc γ RIIB 来调节经典补体途径的激活。它还促进树突状细胞成熟和体液免疫。在心血管疾病中,CRP与氧化的LDL结合,加剧心肌梗死中的组织损伤,并抑制受损血管内皮的修复。