Human Apolipoprotein A1/APOA1 ELISA Kit检测试剂盒
| 规格 | 96 Test |
|---|---|
| 灵敏度 | 91.61pg/ml (50 μl) |
| 标准曲线范围 | 468.75 ~30000 pg/ml |
| 标准曲线梯度 | 7点 |
| 孵育次数 | 2 |
| 可检测样本 | 可溶性物质的液相样品。例如:血清,血浆,细胞培养上清,组织研磨液等。 |
| 加样体积 | 50 μl |
| 类型 | 完全即用型 |
| 操作时长 | 120min |
| pg/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0091 | 0.0093 | 0.0092 | |
| 468.75 | 0.0140 | 0.0140 | 0.0140 | 0.0048 |
| 937.50 | 0.0294 | 0.0286 | 0.0290 | 0.0198 |
| 1875.00 | 0.0781 | 0.0818 | 0.0800 | 0.0708 |
| 3750.00 | 0.2654 | 0.2686 | 0.2670 | 0.2578 |
| 7500.00 | 0.8807 | 0.8709 | 0.8758 | 0.8666 |
| 15000.00 | 2.1960 | 2.2280 | 2.2120 | 2.2028 |
| 30000.00 | 4.4280 | 4.2416 | 4.3348 | 4.3256 |
精度
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 411.0 | 2699.3 | 8436.9 | 609.6 | 2786.8 | 8924.8 |
| Standard Deviation | 24.2 | 110.0 | 280.3 | 21.2 | 15.1 | 427.6 |
| Coefficient of Variation(%) | 5.9 | 4.1 | 3.3 | 3.5 | 0.5 | 4.8 |
板内精密度:3 个已知浓度的样本酶标板内重复测定 22 次,评估酶标板内的精密度。 板间精密度:3 个已知浓度的样本酶标板间重复检测 6 次,评估酶标板间的精密度。
加标回收
健康人血清加入3个不同浓度水平的人Apolipoprotein A1/APOA1,未加入人Apolipoprotein A1/APOA1的血清作为本底,计算回收率。
回收率的范围从93%至120%,平均回收率为107%。
样本值
| Sample Matrix | Sample Evaluated | Range (μg/ml) | Detectable (%) | Mean of Detectable (μg/ml) |
|---|---|---|---|---|
| Serum | 30 | 683.85-2777.78 | 100 | 1035.81 |
应用本试剂盒,检测30份健康志愿者的血清样本。
n.d. =测不到浓度值。样本的浓度值低于灵敏度被认为测不到浓度值。
产品数据表
背景: Apolipoprotein A1/APOA1
已知至少有9种不同的apo蛋白类型。这些apo蛋白起到稳定完整脂蛋白颗粒的作用。人类血清中HDL、LDL和VLDL颗粒的定量测量,常被用于评估个体患冠心病的风险。此外,某些apo蛋白(尤其是A-1和B)的定量免疫学检测,被认为比脂蛋白颗粒的测定方法(尤其是HDL和LDL)更能准确反映冠心病的风险。这些apo蛋白是通过密度梯度超离心法和HPLC从人血浆中提取出来的。